| Number of Output Channels |
R&S® NGP802 / NGP822 |
2 channels |
|
R&S® NGP804 / NGP814 / NGP824 |
4 channels |
| Total Output Power |
R&S® NGP802 / NGP822 |
400 W |
|
R&S® NGP804 / NGP814 / NGP824 |
800 W |
| Output Configuration |
R&S® NGP802 |
2 × 32 V / 20 A |
|
R&S® NGP822 |
2 × 64 V / 10 A |
|
R&S® NGP804 |
4 × 32 V / 20 A |
|
R&S® NGP814 |
2 × 32 V / 20 A; 2 × 64 V / 10 A |
|
R&S® NGP824 |
4 × 64 V / 10 A |
| Maximum Voltage in Serial Operation |
R&S® NGP802 |
64 V |
|
R&S® NGP822 / NGP804 / NGP814 |
128 V |
|
R&S® NGP824 |
250 V |
| Maximum Current in Parallel Operation |
R&S® NGP822 |
20 A |
|
R&S® NGP802 / NGP824 / NGP814 |
40 A |
|
R&S® NGP804 |
80 A |
| Weight |
R&S® NGP802 / NGP822 |
7.5 kg |
|
R&S® NGP804 / NGP814 / NGP824 |
8.0 kg |
| Maximum Output Power per Channel |
200 W |
| Output Channels |
Galvanically isolated and floating |
| Voltage Ripple and Noise |
< 3 mV RMS; < 30 mV peak-to-peak |
| Current Ripple and Noise |
< 3.5 mA RMS |
| Load Recovery Time |
< 400 µs |
| Programming Resolution |
1 mV voltage; 0.5 mA current |
| Readback Resolution |
1 mV voltage; 0.5 mA current |
| Measurement Functions |
Voltage, current, power and energy |
| Protection Functions |
OVP, OCP, OPP, OTP and FuseLink |
| Special Functions |
EasyRamp, output delay, QuickArb and data logging |
| Remote Sensing |
Available for each output at front and rear terminals |
| Remote Control Interfaces |
USB-TMC, USB-CDC and LAN; optional IEEE-488 (GPIB) |
| Optional Control Interfaces |
Digital trigger I/O and analog input |
| Display |
5" WVGA touchscreen, 800 × 480 |
| Operating Temperature |
+5°C to +40°C |
| Power Supply |
100 V to 250 V, 50 Hz to 60 Hz |
| Dimensions (W×H×D) |
362 × 100 × 451 mm |
| Rack Installation |
R&S® ZZA-GE23, 19", 2 HU |